Chapter 4 Performing the Regions of Interest (ROI) Calibration
Creating a Plate Document for the ROI Calibration
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 41
Notes
4
Creating a Plate Document for the ROI Calibration
1. Click (or select FileNew).
2. In the New Document wizard, click to
accept the default parameters.
Note: It is not necessary to name or save the
plate document. The SDS software automatically
saves the ROI data to a set of calibration files on
the computer hard drive.
If a Network communications error is displayed,
click . You can perform the ROI
calibration even if this error is displayed.
Note: This message is displayed if your
computer is not connected to a network. If you
do not plan to connect this computer to a
network, select Do not show me this message
again, then select Never again on this machine.
If an ROI error is displayed, click to
proceed with ROI calibration.
Select
Click
Click
Click