Chapter 6 Performing the Pure Dye Calibration
Creating a Plate Document for the Pure Dye Calibration
72 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Creating a Plate Document for the Pure Dye Calibration
1. In the SDS software, click (or
select FileNew).
2. Configure the New Document dialog box:
a. Select AssayPure Spectra.
b. Select Container96-Well Clear.
c. Select Temp l a teBlank Document.
d. In the Operator field, enter your name.
e. In the Comments field, enter any
information that you want to attach to the
file (such as the plate bar code).
f. Click .
Note: It is not necessary to name or save
the pure dye plate document. The SDS
software automatically saves the pure dye
data to a calibration file on the computer
hard drive.
The Pure Spectra Calibration Manager is
displayed.
Continue with “Preparing Plates and Performing the
Pure Dye Calibration” on page 73.