Chapter 7 Verifying the Instrument Performance
Analyzing the TaqMan
®
RNase P Plate Data
94 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Analyzing the TaqMan
®
RNase P Plate Data
1. In the SDS software, select Analysis
Analysis Settings.
2. In the Analysis Settings dialog box:
a. Select Auto Ct.
b. Click .
3. In the SDS software, click (or select
AnalysisAnalyze).
The SDS software analyzes the run data and
displays the results in the Results tab.
4. Access the results. In the plate document:
a. Select the Results tab.
b. Select the Amplification Plot tab.
Select
b
a
Select
a
b

Table of Contents