Chapter 2 Unpacking and Setting Up the Instrument
Selecting Additional Hardware and Software
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 9
Notes
2
Selecting Additional Hardware and Software
Before beginning the installation, obtain additional software and hardware (if any) you
want to install to the 7300/7500/7500 Fast system.
Choosing Electrical Protective Devices
Applied Biosystems recommends several protective devices to prevent loss of data and
to protect the 7300/7500/7500 Fast system from damage resulting from electrical
hazards.
Power Line
Regulator
Applied Biosystems recommends the use of a 1.5-kVA power line regulator in areas
where the supplied power fluctuates in excess of ±10% of the normal voltage. Power
fluctuations can adversely affect the function of the instrument and the data it produces.
Note: A power line regulator monitors the input current and adjusts the power supplied
to the instrument. It does not protect against power surges or failure.
Uninterruptable
Power Supply
(UPS)
Applied Biosystems recommends the use of a 1.5-kVA uninterruptable power supply
(UPS), especially in areas prone to power failure. Power failures and other events that
abruptly terminate the function of the 7300/7500/7500 Fast system can corrupt data and
possibly damage the system.
IMPORTANT! UPSs have finite battery lives, and, consequently provide power for a
limited time (from 30 minutes to several hours). They are meant to delay the effects of a
power outage, not to serve as replacement power sources. In the event of a power loss,
perform a short-term shutdown of the 7300/7500/7500 Fast system (see page 102) unless
you expect to regain power within the battery life of the UPS.
Surge Protector Applied Biosystems recommends the use of a 10-kVA surge protector (line conditioner)
in areas with frequent electrical storms or near devices that are electrically noisy, such as
refrigerators, air conditioners, or centrifuges. Short-duration, high-voltage power
fluctuations can abruptly terminate the function of and, thereby damage the components
of, the computer and the 7300/7500/7500 Fast instrument.
Note: A dedicated line and ground between the instrument and the building’s main
electrical service can also prevent problems caused by power fluctuations.

Table of Contents

Other manuals for Applied Biosystems 7300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7300 and is the answer not in the manual?

Applied Biosystems 7300 Specifications

General IconGeneral
BrandApplied Biosystems
Model7300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals