Chapter 4 Performing the Regions of Interest (ROI) Calibration
Overview
36 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Overview
Note: You can also use the Real-Time System Setup wizard to perform the procedure in
this chapter. To access the wizard, select Program Files 7300/7500/7500
Fast System Real-Time System Setup Wizard.
Time Required 30 minutes
Materials
Required
When to Perform
the ROI
Calibration
Perform a regions of interest (ROI) calibration:
When installing the system (you must perform all calibrations in this order: ROI
backgroundoptical (7500/7500 Fast systems only)pure dyeinstrument
verification).
Every 6 months, or as often as necessary, depending on instrument use.
After replacing the lamp.
IMPORTANT! After every ROI calibration, you must perform a background calibration,
optical calibration (7500/7500 Fast systems only), pure dye calibration, and instrument
verification.
GR2170
Powder-free
Gloves
Safety
Goggles
ROI Calibration Plate
Centrifuge with plate
adapter
Applied Biosystems
Real-Time PCR System
Spectral Calibration Kit

Table of Contents

Other manuals for Applied Biosystems 7300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7300 and is the answer not in the manual?

Applied Biosystems 7300 Specifications

General IconGeneral
BrandApplied Biosystems
Model7300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals