Chapter 4 Performing the Regions of Interest (ROI) Calibration
Preparing the ROI Calibration Plate
40 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
3. Verify that the liquid in each well of the ROI
plate is at the bottom of the well. If not,
centrifuge the plate again at a higher rpm and for
a longer period of time.
Loading the Plate
1. Press the tray door to open it.
2. Load the ROI calibration plate into the plate
holder in the instrument. Ensure that the plate is
properly aligned in the holder.
Load standard plates
(7300/7500 system)
with the notched A12
position at the
top-right of the tray.
Load Fast plates
(7500 Fast system)
with the notched A1
position at the top-left
of the tray.
3. Close the tray. Apply pressure to the right side of
the tray and at an angle.
Continue with “Creating a Plate Document for the
ROI Calibration” on page 41.
Correct Incorrect
Liquid is at
bottom of well.
Not centrifuged with enough force,
or
Not centrifuged for enough time
Notched
corner
Well A1
12
GR2475
GR2475
7500
keyed corner
A1
A
B
C
D
E
F
G
H
123456789101112
A
1
Press
here

Table of Contents

Other manuals for Applied Biosystems 7300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7300 and is the answer not in the manual?

Applied Biosystems 7300 Specifications

General IconGeneral
BrandApplied Biosystems
Model7300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals