Chapter 4 Performing the Regions of Interest (ROI) Calibration
Preparing the ROI Calibration Plate
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 39
Notes
4
2. Retrieve the spectral calibration kit from the
freezer, then remove the prepared ROI
calibration plate.
3. Return the spectral calibration kit to the freezer.
4. Allow the ROI calibration
plate
to warm to room
temperature (approximately 5 minutes).
IMPORTANT! Do not remove an ROI calibration
plate
from its packaging until you are ready to
run it. The fluorescent dye in the wells of the
plate is photosensitive. Prolonged exposure to
light can diminish the fluorescence signal
strength of the plate.
5. Remove the ROI calibration
plate
from its
packaging. Leave the optical film on the plate.
Do not discard the packaging for the ROI calibration
plate. The plate can be used up to three times if it is
stored in its original packaging sleeve.
1. Standard plates only: Vortex the plate for
5 seconds. Do not vortex Fast plates.
(Remaining steps apply to both standard and Fast
plates.)
2. Briefly centrifuge the ROI calibration plate in a
centrifuge with a plate adapter (<1500xg).
IMPORTANT! The plate must be well mixed and
centrifuged.
Spectral calibration kit
ROI calibration plate
>00:05:00Ambient
00:00:05<1500× g

Table of Contents

Other manuals for Applied Biosystems 7300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7300 and is the answer not in the manual?

Applied Biosystems 7300 Specifications

General IconGeneral
BrandApplied Biosystems
Model7300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals