EasyManuals Logo

Applied Biosystems 7300 User Manual

Applied Biosystems 7300
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Chapter 5 Performing the Background Calibration and Optical Calibration
Overview
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 51
Notes
5
The software then uses the calibration file during subsequent runs to remove the
background fluorescence from the run data.
Background
Fluorescence
Fluorescence data collected by the Applied Biosystems 7300/7500/7500 Fast Real-Time
PCR System includes a fluorescent signal inherent to the system, commonly referred to
as background fluorescence. Background fluorescence is a composite signal found in all
spectral data. This signal consists of fluorescence from several sources, including:
Background electronic signal
Contaminants in the sample block
The plastic consumable (plates and caps)
Guidelines for
Calibration
For a new instrument:
Always start with a new calibration kit.
Make sure the centrifuge you use is clean. Before centrifuging, wipe down the
bucket using a tissue.
Handle the calibration plates with care to prevent contamination. Do not place
plates on a lab bench, which may contaminate the plate. Always put calibration
plates back into their original bags.

Table of Contents

Other manuals for Applied Biosystems 7300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7300 and is the answer not in the manual?

Applied Biosystems 7300 Specifications

General IconGeneral
BrandApplied Biosystems
Model7300
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals