Chapter 3 Auxiliary Operations
IMPORTANT:
Note when setting <Measurement number> to <Assign to ports>:
If you manually set the measurement start number to any number other than “0001” (see step 1 in “2.5.2.
Measuring Samples” on page 2-22), set samples for the first sample rack only in ports of the same number
as the last digit in the measurement start number, or in ports of a higher number than that. Samples set in
ports of a lower number than the last digit will not be measured. This is because measurement operations
are managed so that the port number corresponds to the last digit of the measurement number.
Example: For a measurement start number of “0005”, the first sample rack is processed as follows:
Ports 1 to 4: Measurements are not performed. Ports 5 to 10: Measurements are performed.
If a sample is not set in the port of the same number as the last digit of the measurement start number, the
first measurement number is determined as in the following steps:
1) The last digit of the measurement start number is changed to “0”.
2) The port number of the first detected sample is added to the number obtained in step 1).
Example: Measurement start number: 3005, Ports 5 and 6: No samples, Port 7: Sample set
In this case, the first measurement number is “3007”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals