HA-8180V 1-5
1.1.3 Specifications
Product ADAMS A1C HA-8180V
Configuration Main body, sampler (with sample tube spinning unit) and accessories
Measurement objects Whole blood or hemolysis sample
Column COLUMN UNIT 80
Reagents ELUENT 80A, ELUENT 80B, ELUENT 80CV and HEMOLYSIS WASHING
SOLUTION 80H
Measurement items HbA1
C (stable HbA1C, S-A1C) and HbF
(HbS and HbC can be detected in the Variant mode.)
Measurement ranges *1 HbA1
C: 3 - 20%, 9 - 195 mmol/mol
HbF: 0 - 100%
Guaranteed measurement
ranges *2
HbA1
C: 4 - 16%, 20 - 151 mmol/mol
HbF: 0.3 - 5%
Measurement principle Reversed-phase cation exchange chromatography
Measurement wavelength 420 nm/500 nm (Dual-wavelength colorimetry)
Sample supply Sampler
Resolution 0.1% Ratio, 1 mmol/mol
Processing speed Variant mode: 90 seconds/test
Fast mode: 48 seconds/test
Sample consumption Approximately 14 μL (Whole blood)
Required sample volume Sample tube: Minimum 10 mm away from the bottom
Sample cup: 400 μL or more
Sample container Sample tube: (12.3/15 mm diameter) × (75 to 100 mm length)
Sample cup: 500 μL
Compatible rack type ARKRAY racks
Number of measurement
samples
One-way transportation (factory-set): Maximum 50 samples
Loop transportation: Maximum 100 samples
Column temperature Approximately 40°C
Warm-up time Maximum 30 minutes
Display Large color LCD with backlight
Printer For use with 58-mm width thermal printer paper
Memory capacity 900 measurement results (including calibration results)
External output Serial 1 port (Can be optionally used as an Ethernet port.)
Communication system RS-232C compliant
(Switchable between one-way and two-way communications.)
Transmission speed RS-232C: Selectable from 300, 600, 1200, 2400, 4800, 9600 and 19200 bps
Ethernet: 10BASE-T
Operating environment Temperature: 10 - 30°C; Humidity: 20 - 80% RH (No condensation)
Measurement environment Temperature: 10 - 30°C; Humidity: 20 - 80% RH (No condensation)
Storage environment Temperature: 1 - 30°C; Humidity: 20 - 80% RH (No condensation)
Environment during transport Temperature: 1 - 60°C; Humidity: 20 - 80% RH (No condensation)
(The humidity should not exceed the absolute humidity; 40°C/85% RH.)
Dimensions 530 (W) × 530 (D) × 530 (H) mm
(Not including the hemolysis washing solution bottle)
Weight Main body: Approximately 39 kg, Sampler: Approximately 4 kg
Power requirements 100 - 240 V AC ± 10%, 50/60 Hz
Power input Maximum 300 VA
Sound pressure level Less than 85 dB
Location of use For indoor use only
1.1 Overview