Arkray ADAMS A1C HA-8180V - 3.6.3 Printing a List of Abnormal Results

Arkray ADAMS A1C HA-8180V
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HA-8180V 3-27
3.6.3 Printing a List of Abnormal Results
The list of abnormal results contains the measurement numbers and messages of the results for which “Abnormal
Fraction” appeared on the chromatogram reports. You may use the lists to determine causes of inaccurate measurement
results. Each list contains abnormal results that were obtained on a specific day.
a On the standby screen, select , <6 Print menu> and <3 Abnormal results> in that order.
b Set the setup items.
See the table below.
button: Moves the cursor to the next setup field.
Hyphen button: Moves the cursor through the date in the order
of “day”, “month” and “year”. Or, changes the
option.
c Press .
Printing will start.
The [Print menu] screen will appear again when printing is
complete.
d To return to the standby screen, press twice.
REFERENCE:
Printed report example: See “2.9.4. List of Abnormal Results” on page 2-47.
Messages: See “5.4. Abnormal Result Messages” on page 5-20.
Menu
Abnormal results
(6300) 10-06-2010 10:00
Date (DD-MM-YY)
<10-06-10>
Measurement mode
[Variant ]
PrevItem BS OK Cancel
OK
REFERENCE:
If "None found." appears on the display:
No match is found for the specified date. Press to return to the screen shown in step b.
To stop printing, press or .
Setup item Description
Date (DD-MM-YY) Set a measurement date.
Settable range: 01-01-00 to 31-12-99
Measurement mode Select one or both of the measurement modes for which to print a list. The default is the
current measurement mode.
Variant : Variant mode
Fast : Fast mode
Variant & Fast : Both modes