HA-8180V 1-17
Use the specified RS-232C cross cable (double-shielded cable) to connect an external device to the 9-pin data output
terminal of the instrument. Use of other cables may cause electric shock or fire. For more information, contact your
distributor.
Use the specified Ethernet cable to connect an external device to the Ethernet terminal that is provided when the
optional Ethernet board is installed in the instrument. Use of other cables may cause electric shock or fire. For more
information, contact your distributor.
The START and STOCK terminals are used to connect an optional side sampler. Connecting to other devices may
cause damage to the instrument. For the installation and operating precautions of the side sampler, contact your
distributor. Install the side sampler under the supervision of a service person. Confirm operating precautions with
the service person before use.
Do not disassemble the instrument unless required for installation. Do not modify the instrument. Disassembly and
modification of the instrument may result in exposure to pathogenic microbes or cause fire or damage to the
instrument and consequently lead to personal injury.
If you need to disassemble the instrument after use, wear protective gloves to prevent exposure to pathogenic
microbes.
1.4.2 Unlocking the Instrument
To prevent the instrument from being damaged during transport, various parts are fixed in place before shipping from
the factory. Remove the fixing tape and brackets before installing the instrument. Read “1.4.1. Precautions in
Instrument Installation” on page 1-16 before installing the instrument.
Prepare: Phillips screwdriver
1 Remove the fixing tape.
a Remove the fixing tape from the printer cover, front cover,
maintenance cover and side cover.
REFERENCE:
Keep the removed fixing brackets and screws in the accessory case. These parts should be reused when
transporting the instrument.