HA-8180V 2-22
Chapter 2 Measurement
2.5.2 Measuring Samples
Once samples have been prepared, start measurements.
1 Set the measurement start number (if needed).
On the standby screen, check the measurement start number. Set a different number if necessary. If you do not change
the number, skip to step 2.
a On the standby screen, press .
b Below <Normal measurement samples>, enter a new
measurement start number.
• Range: 0000 to 9999
c Press .
• This saves your entry and will return you to the standby screen.
• The entered number will appear in <Measurement start number>
on the standby screen.
2 Start measurement.
a On the standby screen, check the [Measurement mode] setting.
• To change the mode, see “2.4. Measurement Mode Selection (Variant or Fast)” on page 2-14.
Ready for measurement
10-06-2010 10:00
Measurement start number 0001
Measurement : Variant
Column 0000
ABCH
Menu Meas No. Priming Cal
Meas No.
Measurement start number
10-06-2010 10:00
Normal measurement samples
<0300>
HbA1c control
<0001>
PrevItem BS OK Cancel
This instrument stirs samples before measuring them. Due to a possible risk of injury, do not touch the
spinning unit or sample container, or insert anything in-between the STAT port cover and front cover while the
sample tube is spinning. Also, check that the STAT port cover is attached correctly before starting
measurement.