HA-8180V 2-49
2.9.7 Calibration Results Report
The instrument automatically prints calibration results report in the following cases:
When calibration results have been obtained after calibration
Before starting measurement of samples
This report can be reprinted if needed (see “3.6.4. Printing Calibration Result Report” on page 3-28).
Example: Automatic calibration
Example: Calibration coefficients setting
HbA1c calibration result
Var.
10-06-2010
10:00
Calibration method
[
Automatic
]
Lot
CA89C05
Expiration date
2011-05
Calibration error range (%)
<3.0>
Standard value (mmol/mol)
L
<35>
H
< 89>
Results
34.8
87.2
34.8
87.5
34.8
87.5
Mean
34.8
87.4
Calibration coefficients
a<1.0266>
b<- 0.7>
Conversion to NGSP: Y = a X + b
a<0.0915>
b<+2.15>
User-specified coefficients
HbA1c
a<1.00>
b<+ 0.0>
HbF
a<1.00>
b<+0.00>
Lot No. Expiry
Elu. 80A 0A1101 2011-01
Elu. 80B 0C1112 2011-03
Elu. 80CV 0D3331 2011-04
Wash 80H 9L1121 2010-12
HbA1
C
measurement results (mmol/mol)
Calibration date and time
Selected measurement mode: Var.(Variant) or Fast
Calibration method (automatic calibration)
Lot number of a calibrator
(when using calibration information barcode)
Expiration date of a calibrator
(when using calibration information barcode)
Calibration error range
Calibration standard values (mmol/mol)
Mean values of HbA1
C
measurement results (mmol/mol)
Calibration coefficients of HbA1
C
Coefficients of the conversion formula to NGSP
User-specified coefficients for HbA1
C
User-specified coefficients for HbF
Reagent information
Hyphens (---) appear if no information has been set.
HbA1c calibration result
Var.
10-06-2010
10:00
Calibration method
[
Calibration coefficients setting
]
Calibration coefficients
a<1.0266>
b<- 0.7>
Conversion to NGSP: Y = a X + b
a<0.0915>
b<+2.15>
User-specified coefficients
HbA1c
a<1.00>
b<+ 0.0>
HbF
a<1.00>
b<+0.00>
Calibration date and time
Selected measurement mode: Var.(Variant) or Fast
Calibration method (calibration coefficients setting)
Calibration coefficients of HbA1
C
Coefficients of the conversion formula to NGSP
User-specified coefficients for HbA1
C
User-specified coefficients for HbF
2.9 Displayed and Printed Reports