HA-8180V 6-8
Chapter 6 Appendix
Example: Diagnosis results
6.1.6 Reagent Information Report
Reagent information can be printed out as follows:
At the beginning of batches
On individual measurement result reports
Reagent information is not printed as a default. If needed, change the printer settings (see “3.5.2. Setting Up the
Printer” on page 3-23).
Example
HbA1c reproducibility test
Fast
10-06-2010 10:00
HbA1c
HbF
1 33 0.8
2 33 0.8
3 33 0.7
4 33 0.8
5 34 0.8
:
:
10 33 0.8
Mean
33.1
0.79
R
1
0.1
S.D.
0.3
0.03
C.V.
0.9
Print date
Current measurement mode
Result obtained by each measurement
Mean value of measurement results
Difference between maximum and minimum measurement
results
Standard deviation
Coefficient of variation
Lot No. Expiry
Elu. 80A 0A1101 2011-01
Elu. 80B 0C1112 2011-03
Elu. 80CV 0D3331 2011-04
Wash 80H 9L1121 2010-12
Calib. CA89C05 2011-05
Hyphens (---) appear if no information has been set.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals