HA-8180V 2-11
2.2.5 Samples
Anticoagulant for whole blood
Use one of the following anticoagulants: heparin, EDTA-2Na, EDTA-2K, EDTA-3K or NaF.
Never measure samples that contain iodoacetic acid as an anticoagulant in order to prevent degradation of the column.
When storing whole blood samples:
Refrigerate whole blood samples at a temperature between 2°C and 8°C. Samples can be stored in stable condition for
3 to 4 days before use.
TAKE THE UTMOST CARE WHEN HANDLING BLOOD. This instrument uses blood as
sample. Blood may be contaminated by pathogenic microbes that can cause infectious
diseases. Improper handling of blood may cause infection to the user or other individuals by
pathogenic microbes.
Discard used samples in accordance with local regulations for biohazardous waste.
IMPORTANT:
When using hemolysis samples:
If left at room temperature, hemolysis samples gradually degenerate, which will affect measurement.
Promptly measure hemolysis samples without leaving them at room temperature for any extended periods
of time.
Hemoglobin concentration of hemolysis samples
Prepare hemolysis samples so that the hemoglobin concentration is between 75 mg/dL and 225 mg/dL
(standard: 150 mg/dL). If the concentration does not fall within this range, an error may occur or the
reproducibility may become poor. If this happens, change the dilution ratio appropriately to adjust the
concentration before performing a measurement again. (This instrument dilutes whole blood 101 times
before measuring it.)
Samples without plasma
Sample racks for whole blood cannot be used to measure samples from which plasma has been removed.
With the whole blood racks, samples without plasma cannot be diluted to the proper hemoglobin
concentration and consequently produces inaccurate measurement results. To measure such samples,
dilute them with Diluent 80 and measure as hemolysis samples.
2.2 Measurement Precautions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals