HA-8180V 3-14
Chapter 3 Auxiliary Operations
Setup item Description (default bolded)
Control expected
values (mmol/mol)
Set control expected values of the controls you are using. These values are written on the
package insert of the controls.
L: Low solution 0 to 99 mmol/mol (0 mmol/mol)
H: High solution 0 to 200 mmol/mol (0 mmol/mol)
Control
measurement count
Set the number of times for measuring the control in a single sample container. Up to 3
measurements can be performed per sample container, so you can reduce the volume of
controls used for control measurements.
Range: 1 to 3 times (3 times)
Control error range
(mmol/mol)
Set the error detection range used in control measurements. The instrument issues a warning
(W:011 or W:071) if the difference between the obtained control measurement result and
<Control expected values> exceeds the values set here.
L: Low solution 0 to 99 mmol/mol (3 mmol/mol)
H: High solution 0 to 99 mmol/mol (4 mmol/mol)
Operation if control
error occurs
Select the action to take place when the difference between the obtained control
measurement result and <Control expected values> exceeds <Control error range>.
Stop measurement: Issues W:071 and stops measurements.
Issue warning: Issues W:011, but continues measurements.
No action: Continues measurements without issuing a warning.
Pressure unit Select the pressure unit for the column.
Options: kg/cm2, MPa
STD. solution
measurement count
Set the number of times to measure each of the Low and High solutions for automatic
calibration.
Range: 1 to 3 times (3 times)
Calibration error
range (%)
Set the error detection range used in automatic calibration.
Range: 0.0 to 9.9% (3.0%)
Column degradation Set whether or not to notify the user by a message that the column requires replacement
when the column measurement counter passes a preset number.
Misread barcodes Set the number of misread barcode errors in a batch at which the instrument issues W:081
and stops measurement. If set to “0”, measurement does not stop regardless of times the
error occurs.
Range: 0 to 150 times (0 time)
Sample tube spin
failures
Set the number of sample tube spin failures in a batch at which the instrument issues W:082
and stops measurement. If set to “0”, measurement does not stop regardless of times the
failure occurs.
Range: 0 to 150 times (0 time)