HA-8180V 4-52
Chapter 4 Maintenance
3 Turn on the power.
a Plug the power cord into an outlet.
b Press the “–” side of the main power switch.
c Press the standby switch to turn on the power.
4 Set the date and time.
a Check the date and time on the display, and adjust them if necessary.
See “3.5.1. Setting the Date and Time” on page 3-22.
5 Perform priming.
a Perform priming.
See step 1 in “1.5.2. Installing the Column” on page 1-31.
6 Install the column.
a Install the column.
See “4.2.4. Replacing the Column” on page 4-11.
In step 3, remove the “dummy column” instead of the “old column”.
In step 9-b, do not reset the column measurement counter.