HA-8180V 5-19
T:801 - T:804
T:807 - T:810
Two-way communication
Cause The communication cable of the external device is disconnected or improperly connected.
Remedy a Press to clear the trouble.
b Properly connect the communication cable.
c If the trouble persists, turn off the power and contact your distributor.
T:8 11 No matching ID
Cause The host computer transmitted an abnormal measurement command.
Remedy a Press to clear the trouble.
b Check if the host computer correctly replies to an inquiry from the instrument.
T:820 - T:822 Ethernet communication
Cause The Ethernet board did not operate properly.
Remedy a Press to clear the trouble.
b Turn off the power and contact your distributor.
T:9 99 Other trouble
Cause Other trouble occurred.
Remedy Take notes of what appears on the screen, and contact your distributor.
5.3 If Trouble Occurs