Chapter 2 Measurement
b Set the sample tubes in the ports of the sample racks.
When preparing sample cups with sample:
Use the whole blood pair rack (W PAIR) or hemolysis pair rack (H PAIR).
a Set sample tubes with barcode labels in the odd-numbered ports.
b Set sample cups with sample in the even-numbered ports.
Example: The barcode of port 1 is assigned to the sample in
port 2.
Rubber cushions
Sample tube
Fit the bottom of the sample tubes into the
bottom of the rubber cushions so that the tubes
stand straight. If the tubes are tilted, they may
cause damage to the piercing nozzle.
Barcode label
Rear of sample rack
For sample tubes labeled with barcode:To enable the
internal barcode reader to read barcodes successfully,
labels on the tubes must be facing the rear of the sample
rack.
Sample tube with barcode label
Sample cup with sample
12345678910
Do not set whole blood samples in the even-numbered ports of the hemolysis pair rack to avoid seriously
degrading the column. If whole blood is measured in these ports, it is recommended to replace the column
with a new one.
Set barcoded sample tubes in sample racks with the barcode labels facing the rear of the sample rack.