Chapter 2 Measurement
For sample cups with control (when barcodes are set):
Odd-numbered ports: Sample tubes with barcode labels
[12.3 mm diameter]
Use the adapter on the rack.
[15 mm diameter]
Remove the adapter.
Even-numbered ports: Sample cups with control
Example: The barcode of port 1 is assigned to the control in port 2.
For sample tubes with control:
Odd-numbered ports: Sample tube with control (attach barcode
labels if needed)
[12.3 mm diameter]
Use the adapter on the rack.
[15 mm diameter]
Remove the adapter.
Even-numbered ports: Leave the ports empty.
4 Load the hemolysis control rack onto the sampler.
a Load the hemolysis control rack onto the sampler.
See step 4 in “2.5.1. Preparing Samples” on page 2-21.
NOTE:
To perform normal measurements following control measurement, be sure to set the samples in either normal
rack, whole blood pair rack or hemolysis pair rack. Do not set whole blood samples in the unused ports of the
hemolysis control rack to avoid seriously degrading the column. If whole blood is measured with the hemolysis
control rack, it is recommended to replace the column with a new one.
IMPORTANT:
Do not set a sample cup in the port next to a sample tube with control. The sample tube will be recognized for
barcode reading and the control in the tube will not be measured. Also, the barcode of the sample tube will be
assigned to the sample in the sample cup.
NOTE:
Set sample tubes with barcode labels in the hemolysis control rack with the barcode labels facing the rear of
the rack.
REFERENCE:
If sample rack transportation is set to “loop transportation”:
Load the empty start rack and then the hemolysis control rack onto the sampler.
To perform normal measurements following control measurement:
Load the hemolysis control rack and then normal racks onto the sampler.
Empty sample tube with barcode label
Sample cup with control
12345678910
Sample tube with control
* Barcode labels can be attached.
Empty
12345678910

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals