15. Patient Data Card
Data Item Data Range Increment Units
43
Max Na Buttons per
Treatment
[0, 7] (range can be reduced
based on calibration set
maximum [0, 7])
1 N/A
44 DPSND Tidal Volume
[20, 120] (range can be
reduced based on calibration
set minimum [0, 120] and set
maximum [minimum, 120])
1 mL
45 DPSND AV Ratio
[0.5, 2.0] (Represents ratio
[1.0:X.X])
0.1 N/A
46 DPSND Needle Volume [0.0, 10.0] 0.1 mL
47 HDF Target Volume
[0.0, 18.0] (0 indicates option
not being used for treatment,
and range can be reduced
based on a maximum equal
to prescribed time maximum
HDF rate
0.1 L
48 OLHDF Infusate Volume
[0.0, 60.0], 0 = disabled
(range can be reduced based
on calibration set minimum
[0.1, 3.0] and set maximum
[minimum, 60.0]. Range can
be further reduced based on
a maximum equal to
prescribed time maximum
OLHDF rate)
0.1 L
49 Treatment Mode
{Standard, Batch HDF,
OLHDF, DPSND, SPSND},
where Standard=0, Batch
HDF=1, etc. * See Note
below this table
N/A N/A
50
Blood Pressure Monitor
Required
{No, Yes} N/A N/A
51
UF Profiling Option
Required
{No, Yes} N/A N/A
52
Sodium/Bicarb Profiling
Option Required
{No, Yes} N/A N/A
53 Heparin Pump Required {No, Yes} N/A N/A
54 Hemavision Required {No, Yes} N/A N/A
55 Sodium Button Required {No, Yes} N/A N/A
56
DPSND Option
Required
{No, Yes} N/A N/A
57 HDF Option Required {No, Yes} N/A N/A
58 OLHDF Option Required {No, Yes} N/A N/A
59 SPSND Option Required {No, Yes} N/A N/A
NOTE
In Line 49 of the table above, the Treatment Mode setting must
also be compatible with the Option Required settings, as shown in
Table 15-2.
157-1278-922 Rev A 15-7
January 2004