MX08 Multichannel Weighing System, Technical Manual, Rev. 2.5 Page 99 of 144
Answer : 03, 0E, 04, 22, 00, 04, BF, 38, 00, 04, BF, 38, 00, 00, 00, 00,
Status : 0422 hex
Indicated : 0004BF38 hex (311096 dec)
Gross : 0004BF38 hex (311096 dec)
Tare : 00000000 hex (0 dec)
Read inputs of MX08 IO [0] at register 41071.
Request : 03, 04, 2E, 00, 01
Answer : 03, 02, 05, 0D
Inputs : 0D hex (1011 binary)
Outputs : 05 hex (0101 binary)
Set all outputs of MX08 IO [0] at register 41071.
Request : 10, 04, 2E, 00, 01, 02, 00, 0F
Answer : 10, 04, 2E, 00, 01
Outputs activated.
Zeroing MX08 AD [0] at register 41062.
Request : 10, 04, 25, 00, 01, 02, 00, 01
Answer : 10, 04, 25, 00, 01
MX08 AD [0] is zeroed.
Zeroing MX08 AD [1] at register 41063.
Request : 10, 04, 26, 00, 01, 02, 00, 01
Answer : 10, 04, 26, 00, 01
MX08 AD [1] is zeroed.
Taring MX08 AD [0] at register 41062.
Request : 10, 04, 25, 00, 01, 02, 00, 02
Answer : 10, 04, 25, 00, 01
MX08 AD [0] is tared.
Taring MX08 AD [1] at register 41063.
Request : 10, 04, 26, 00, 01, 02, 00, 02
Answer : 10, 04, 26, 00, 01
MX08 AD [1] is tared.
Zero Calibration of MX08 AD [0].
Request : 03, 04, 66, 00, 01 (Read status; it must be in ready status)
Answer : 03, 02, 00, 01 (MX08 AD [0] is in ready status; zero calibration can be
performed)
Request : 10, 04, 63, 00, 01, 02, 00, BC (Zero calibration command)
Answer : 10, 00, 6D, 00, 01
Request : 03, 04, 66, 00, 01 (Read status and wait to become in ready status)
Answer : 03, 02, 00, 01 (If it changed “zeroing” to “ready” status, zero calibration is
performed successfully.)
Span Calibration of MX08 AD [0].
Request : 03, 04, 66, 00, 01 (Read status; it must be in ready status)
Answer : 03, 02, 00, 01 (MX08 AD [0] is in ready status; span calibration can be
performed)
Request : 10, 04, 63, 00, 03, 06, 00, DC, 00, 00, 0B, B8
(Span calibration command with 3000 (0x0BB8 hex) span value)
Answer : 10, 00, 6D, 00, 03
Request : 03, 04, 66, 00, 01 (Read status; it must be in ready status)
Answer : 03, 02, 00, 01 (If it changed “span calb” to “ready” status, span calibration is
performed successfully.)