BD BACTEC™ FX Instrument User’s Manual
160
09
One or more
stations has had a
measurement
failure or has had a
vial presence
switch failure. Open
drawer to resolve
error(s).
Instrument has detected a
condition that could represent
a measurement failure or
partially seated vial.
Alert clears when the instrument
detects that the failure no longer exists.
Make sure all vials are fully seated in
the stations.
If alert does not clear, block the station
and contact BD.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
11
Printer offline. The printer’s pending queue
is temporarily full.
The printer is turned off.
The print cable is unplugged.
The connection to the hub is
disconnected.
The print server is turned off.
The hub power supply is
disconnected.
The instrument is offline in a
Workgroup.
Check all conditions listed at left. Any
queued reports should print when error
is corrected.
13
Database
Reinitialized.
Set during startup when the
instrument detects corruption
in the database.
Alert clears when the instrument
reinitializes the database and
eliminates the corruption.
14
Time offset (SESC)
will inhibit the
movement of vials
between
instruments.
Contact BD.
A time mismatch has
occurred between multiple
instruments.
This error is set and cleared
automatically. Contact your local BD
representative if error persists/ recurs.
15
Contains a vial
which may have a
reading gap due to
an invalid time
offset (SESC).
Consult manual.
Most likely cause is that a vial
was moved between
instruments with mismatched
time offsets.
Any vials with reading gaps are
marked as Affected Vials. Refer to
Section
7.3.3 for instructions on
Affected Vials.
16
EpiCenter
Communications
failure.
This alert is detected when
BD EpiCenter is configured
and cannot be reached.
Alert clears itself when communication
with BD EpiCenter is reestablished.
Error
No.
Message Possible Cause(s) Corrective Action(s)