BD BACTEC™ FX Instrument User’s Manual
180
9 Select Exit to return to the Status display.
10 If there are any negative vials in the drawer, select Remove Negatives on the Status display.
11 Remove all negative vials using the negative removal activity (Section 4.9).
12 When all negative vials are removed, the instrument beeps 3 times to indicate that this activity
is complete.
13 Select Exit to return to the Status display.
14 If there are any anonymous vials in the drawer, select Identify Anonymous on the Status
display.
15 Identify all anonymous vials using the ID Anonymous activity (Section 4.5). If there are no
available stations in the current drawer, select Save to save the identification. Use Vial Entry to
reenter the vials in a drawer with available stations when you are done identifying anonymous
vials. (Perform Step 26 for these vials if you are going to return them to the instrument
immediately.)
16 When all anonymous vials are identified, the instrument beeps 3 times to indicate that this
activity is complete.
17 Close the drawer, wait a moment, and reopen the drawer.
18 Message WE31 appears again.
19 Select Later in the message box.
20 Observe the Status display. If any of the vials identified in steps 14 through 16 have gone out
of protocol, repeat steps 10 through 13.
21 Close the drawer, wait a moment, and reopen the drawer.
22 Message WE31 appears again.
23 Select Now in the message box. When you select Now, the display shown in Figure 7-1
appears.
24 Remove vials with the RED (solid) LEDs from their stations.
25 Because you earlier removed all positive and negative vials, what remains are ongoing vials.
26 Before returning the ongoing vials to good stations, check the vials’ sequence numbers
against the Affected Vial Report that you printed. For optimal recovery, subculture all affected
vials before returning to them to usable stations.
27 Use the Vial Entry activity to return the vials to good stations (Section 4.5). Vials should be
returned within 5 hours.
28 Block all unusable stations (Section 6.2.2.1).