2 – Installation
27
Accessions can always be entered in the Culture – Vial display regardless of whether
accession barcoding is enabled or disabled.
In a BD EpiCenter configuration, enabling Accession Barcoding at one instrument enables the
feature for all instruments.
Select whether Batch Negative Removal is enabled (checked) or disabled (unchecked) by
toggling the white box to check on or off. The default setting is enabled (checked). When Batch
Negative Removal is enabled, you can remove all final (out-of-protocol) negative vials without
scanning the individual labels during the Negative (vial) Removal activity. When batch removal
is disabled, you must scan the vial sequence barcode for each negative vial that is removed.
In a BD EpiCenter configuration, enabling Batch Negative Removal at one instrument enables
the feature for all instruments.
Select whether Show Related Vials is enabled (checked) or disabled (unchecked) by toggling
the white box to check on or off. The default setting is disabled (unchecked). When Show
Related Vials is enabled, all vials in the same drawer that are related to the current positive vial
being removed (i.e., vials with the same accession number) are flagged by station indicators
that correspond to their status (Ongoing, Negative, etc.).
In a BD EpiCenter configuration, enabling Show Related Vials at one instrument enables the
feature for all instruments.
Language/Locale Window:
Select the language for screen displays and instrument reports. The default setting is English.
To see the available selections, select the Down Arrow button to the right of the field. You can
choose from the following language selections:
Deutsch
English
Español
Français
Italiano
Japanese
In a BD EpiCenter configuration, selecting a language at one instrument selects that language
for all instruments.
Batch Negative Removal Enable/Disable
Show Related Vials Enable/Disable
Language