6.2 Software Functions and System Configuration
64
Moisture %
Moisture reading when the radiometric bulk flow measurement is
configured accordingly and a moisture meter is connected (only in
operating mode with moisture compensation).
Dry totalizer (t or kg)
Display of dry weight in the selected unit (only in operating mode
with moisture compensation).
Raw data
By pushing the <more> button, the raw data of the connected de-
tectors can be displayed live. The following raw data can be read
for each detector:
HV
Shows the actual high voltage value at the detector.
HV-auto: Automatic HV control is enabled. This is the normal mode
for the HV.
HV test: A manual fixed value has been set in the Service menu.
This setting should only be used for a test.
cps
Display of the unaveraged live count rates of this detector.
Det. Temp.
Display of the detector temperature.
Service Menu
See Service Functions on page 95.