CB (E6.1) 07/2017 Page 23/145
In order to avoid contamination, never place the chamber directly on the floor.
CAUTION
Danger of overheating.
Damage to the chamber.
Do NOT set up chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Do not install or operate the chamber in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Ambient conditions
Permissible ambient temperature range for operation: +18 °C / 64.4 °F to +30 °C / 86 °F
At elevated ambient temperature values, fluctuations in temperature can occur.
Ideal ambient temperature: at least 7 °C / 12.6 °F below the intended working temperature. E.g.,
working temperature 37 °C / 98.6 °F = ambient temperature 30 °C / 86 °F and lower. In the event of
working temperatures of less than 7 °C / 12.6 °F above the ambient temperature, the setpoint can be
exceeded.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of 22 ±3 °C / 71.6 ±5.4 °F to which the specified technical data relates. For other
ambient conditions, deviations from the indicated data are possible.
Avoid direct solar radiation on the chamber.
Permissible ambient humidity: 70 % r.H. max., non-condensing.
Installation height: max. 2000 m / 6561.7 ft above sea level.
Wall distances: rear 100 mm / 3.94 in, sides 50 mm / 1.97 in.