Balance System SD (with v4.X software) 115
13. Electromagnetic Compatibility
Conformance to Standards
This equipment conforms to the following safety standards:
Table 13.1. Safety Standards Conformance Table
Standard
Edition and/or date
IEC60601-1-2
First Edition, 2007
Accompanying EMC Documents
This medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Use of accessories, transducers, and cables other than those specified, with the
exception of accessories, transducers, and cables sold by the manufacturer of this
equipment, as replacement parts for internal and external components, may result in
increased emissions or decreased immunity of the equipment.
The Balance System SD should not be used adjacent to or stacked with other equipment.
If the Balance System SD is used while positioned adjacent to other equipment, it should
be observed to verify normal operation in the configuration in which it will be used.
List of Cable Accessories
The list in Table includes all accessory cables supplied with the Balance System SD for which the
manufacturer of this equipment claims compliance to EN 60601-1-2 when used with the Balance
System SD.
Table 13.2 Balance System SD Cables
Cable Description
Part No.
Cable Length
USB Printer Cable
Biodex # C12086
15ft