BCS-8xx series installation manual
20
2.1.3 Ventilation Requirements
The BCS-8xx units must be installed at least 0.8 m from the wall.
Fig. 21: Ventilation requirements for BCSC cabinets.
2.1.4 Electrical Plugs
Caution:
We strongly recommend to use National Plug conform to national or IEC standard to con-
nect the power cable to the Electrical installation. A permanent connection is prohibited.
Following some examples of possible national three phase plugs/receptacles for the BCSC
cabinet.
BCSC Cabinet with North American PDU (US version)
Three-Phases Plug/Receptacle
Countries
BCSC-24U
BCSC-38U
USA
or
CAN
or
JP
NEMA 15-30 (3W+PE)
or
NEMA L15-30 (3W+PE)
Or
NEMA L21-30 (3W+N+PE)
in this case the Neutral "N" is
not used.
NEMA 15-50 (3W+PE)
or
NEMA L15-50 (3W+PE)
Or
NEMA L21-50 (3W+N+PE)
in this case the Neutral "N" is not
used.