BCS-8xx series installation manual
62
8.3 Environmental
- Indoor use
- Operating Temperature: 5°C to +40°C Indoor use (max power derating starts at
23°C and is dependant of cell cable length see power contour plot in ch. 7.4)
- Storage Temperature: 0°C to +50°C
- Pollution degree: 2 (normally only non-conductive pollution occurs)
- Altitude: <2000 m above sea level
- Humidity: 10% to 80% non-condensing for temperatures up to 31 °C decreasing
linearly to 50 % relative humidity at 40 °C
- MAINS supply voltage: fluctuations up to ±10 % of the nominal voltage
- Transient overvoltage: typically present on the MAINS supply.
- Case protection: IP20 or better
- Warm-up: 1 hour to rated accuracy
- Cooling: Internal DC Fans
- Vibration: not specified
- Choke: not specified
Safety: complies with EN61010-1.
EMC: complies with EN61326.
RoHS: complies with 2002/95/EC