7.4 Preparation before cleaning
Before the first automatic cleaning/disinfection: attach ECCOS fixa-
tions in a suitable sieve basket.
Insert the products in the correct position into the ECCOS fixations, see
Fig. B.
Kirschner wire attachment GB641R
Turn adjustment sleeve 20 to setting for largest Kirschner wire diam-
eter.
7.5 Cleaning/disinfection
7.5.1 Product-specific safety instructions for the reprocessing pro-
cedure
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
Use cleaning and disinfecting agents according
to the manufacturer’s instructions which
are approved for plastic material and high-
grade steel,
do not attack softeners (e.g. in silicone).
Do not use cleaning agents that contain ace-
tone.
Observe specifications regarding concentration,
temperature and exposure time.
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
Do not exceed the maximum temperature of
96 °C during thermal disinfection with FD
water.
Dry the product for at least 10 minutes at a
maximum of 120 °C.