4
Aesculap
®
Acculan 4
Attachments
2.2 Main functions and design characteristics 2.3 Indications
The type and area of application depend on the tool selected.
2.4 Absolute contraindications
The product is not licensed for use on the central nervous system or cen-
tral circulatory system.
2.5 Relative contraindications
The safe and effective use of the product greatly depends on influences
which can only be controlled by the user. Therefore the specifications pro-
vided represent framework conditions only.
Clinically successful use of the product is dependent on the knowledge
and experience of the surgeon. The surgeon must decide which structures
it is sensible to treat and take into account the safety and warning infor-
mation contained in these instructions for use.
3. Safe handling
CAUTION
Federal law restricts this device to sale by, or on order of a physician!
General risk factors associated with surgical procedures are not
described in this documentation.
It is the operating surgeon's responsibility to ensure that the surgical
procedure is performed correctly.
The operating surgeon must have a thorough understanding of both
the hands-on and conceptual aspects of the established operating
techniques.
Remove the transport packaging and clean the new product, either
manually or mechanically, prior to its initial sterilization.
Prior to use, check that the product is in good working order.
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability:
Use the product only according to these instructions for use.
Follow the safety and maintenance instructions.
Only combine Aesculap products with each other.
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
Speed GB667R
GB668R
GB669R
GB670R
0 min
-1
to max. 250 min
-1
GB657R
0 min
-1
to max. 370 min
-1
GB645R
0 min
-1
to max. 900 min
-1
GB620R
GB621R
GB663R
GB664R
GB665R
0 min
-1
to max. 1 000 min
-1
GB623R
GB639R
GB641R
0 min
-1
to max. 1 250 min
-1
GB660R
0 min
-1
to max. 17 000 min
-1
Rotational
direction
Right and left rotation, oscillation
Operating
mode
Operation with non-periodic load and speed
changes (type S9 pursuant to
IEC EN 60034-1)
Right/left rotation:
60 second application, 60 second pause
20 repetitions
30 min cooling time
Max. Temperature 48 °C
Milling (clockwise/counterclockwise):
30 second application, 30 second pause
8 repetitions
30 min cooling time
Max. Temperature 48 °C
Drilling (oscillation):
60 second application, 60 second pause
4 repetitions
30 min cooling time
Max. Temperature 48 °C
Saw mode with GB660R:
30 second application, 60 second pause
3 repetitions
30 min cooling time
Max. Temperature 48 °C
WARNING
Risk of injury and material damage due to inappro-
priate use of the product!
Use the product only in accordance with the
intended purpose.
WARNING
Risk of injury and damage to property due to
improper handling of the product!
Follow the instructions for use of all products
used.