Attachments
9. Troubleshooting list
► Have defective products repaired by Aesculap Technical Service, see
Technical Service.
Issue Cause Detection Remedy
Product becomes too hot Excessive use Attachment very hot Follow operating instructions (operating
mode).
Defective attachment drive/ball
bearings
Attachment very hot Follow operating instructions (preparation,
care).
Preventative: lubricate attachment before
every sterilization.
Have the manufacturer repair the attachment.
Fall damage, product defective Attachment very hot Have the attachment repaired by the manu-
facturer
Blunt tool Heating of tool and product Change tool.
Insufficient power Product is operating in left rota-
tion mode
Serrated tool run in counterclock-
wise mode
Use serrated tools only in clockwise mode
Attachment defective Attachment very hot Follow operating instructions (preparation,
care).
Preventative: lubricate attachment before
every sterilization.
Follow operating instructions (operating
mode).
Have the attachment repaired by the manu-
facturer
Blunt tool Tool blades worn out Change tool.
Loud running noise Defective attachment drive/ball
bearings
Loud, perceptible noise during
operation
Follow operating instructions (preparation,
care).
Preventative: lubricate attachment before
every sterilization.
Have the manufacturer repair the attachment.
Attachment does not
attach or detach.
Attachment not compatible Attachment doesn't engage Use the compatible attachment for drill.
Connection on attachment
deformed/defective
Attachment is difficult to couple
or cannot be coupled
Have the manufacturer repair the attachment.
Coupling on drill deformed/defec-
tive
Attachment is difficult to couple
or cannot be coupled
Have the manufacturer repair the drill.
Sleeve on drill hard to move Attachment doesn't engage Turn and hold the drill sleeve, then attach the
attachment. Turn the sleeve back if necessary.
Preventative: lubricate the sleeve on the drill
before each sterilization.