7.11 Sterilization for the US market
■ Aesculap advises against sterilizing the device by flash sterilization or
chemical sterilization.
■ Sterilization may be accomplished by a standard prevacuum cycle in a
steam autoclave.
To achieve a sterility assurance level of 10
-6
, Aesculap recommends the
following parameters:
*Aesculap has validated the above sterilization cycle and has the data on
file. The validation was accomplished in an Aesculap sterile container
cleared by FDA for the sterilization and storage of these products. Other
sterilization cycles may also be suitable, however individuals or hospitals
not using the recommended method are advised to validate any alterna-
tive method using appropriate laboratory techniques. Use an FDA cleared
accessory to maintain sterility after processing, such as a wrap, pouch, etc.
7.12 Storage
► Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8. Maintenance
To ensure reliable operation, the product must be maintained at least once
a year.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*
Sterilization
method
Temp. Time Minimum drying time
Prevacuum 270 °F/275 °F 4 min 20 min