13
7.7 Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
7.7.1 Manual pre-cleaning with a brush
D–W: Drinking water
RT: Room temperature
► Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
► For Kirschner wire attachment GB641R: Set the adjustment sleeve 20
to the greatest Kirschner wire diameter.
Phase I
► Mobilize non-rigid components during cleaning.
► Thoroughly clean the product under running water.
Phase II
► Mobilize non-rigid components during cleaning.
► Brush cannulation with cleaning brush TA011944 and difficult to
access surfaces with a suitable cleaning brush made of plastic for at
least 1 min.
► After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available on Aesculap Extranet at
https://extranet.bbraun.com).
Phase Step T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemistry/Note
I Rinsing RT (cold) - - D–W Until visually clean
II Brushes RT (cold) - - D–W Until visually clean