14
Aesculap
®
Acculan 4
Attachments
7.7.2 Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
D-W: Drinking water
FD–W: Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
For Kirschner wire attachment GB641R: Set the adjustment sleeve 20
to the greatest Kirschner wire diameter.
Insert the product in its proper position in the ECCOS fixation.
Then connect the interior rinse device to the ECCOS fixation and con-
nect to the rinse connector of the cleaning/disinfection device/rinse
cart.
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.8 Inspection, maintenance and checks
Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Spray product after every cleaning and disinfection with Acculan 4 oil
spray adapter 23 GB600480 (black) for approx. 2 sec. with Aesculap
STERILIT Power Systems oil spray GB600, see Fig. C.
Note
Aesculap recommends occasional spraying of moving parts (such as knobs,
couplings, cover caps) with the Aesculap STERILIT-Power-Systems oil
spray.
Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration.
Set aside the product if it is damaged.
7.9 Packaging
Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for Aesculap-
ECCOS holder system).
Insert the products in the correct position into the ECCOS fixations,
see Fig. B.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.
7.10 Steam sterilization
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
Use a validated sterilization method:
Steam sterilization using fractional vacuum process
Steam sterilizer DIN EN 285 and validated pursuant to
DIN EN ISO 17665
Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing multiple products in one steam sterilizer:
Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.
Phase Step T
[°C/°F]
t
[min]
Water
quality
Chemistry/Note
IPre-rinse <25/77 3 D–W -
II Cleaning 55/131 10 FD–W
Concentrate, alkaline:
–pH ~ 13
<5 % anionic surfactant
0.5 % working solution
–pH ~ 11*
III Intermediate rinse >10/50 1 FD–W -
IV Thermal disinfection 90/194 5 FD–W -
VDrying - - - mind. 10 min at max. 120 °C