Technical Data
70 H148850_3_003
14.2 Electrical Characteristics
14.2.1 Power Requirements
Power Supply Input:
AC Voltage, Frequency, and Current
100-240 V ~
47-63 Hz
1.07 A
Installation Category: CAT II (Line voltage in appliance and to
wall outlet)
Output:
24 V DC, 3.33 A
Sample Changer Input:
DC Voltage and Current
24 V
3.33 A
Installation Category: CAT I (Mains isolated)
Use only with the provided power supply.
Sample Stage Power
Supply
Input:
AC Voltage, Frequency, and Current
100-240 V ~
50-60 Hz
0.5 A
Installation Category: CAT I (Mains isolated)
Output:
24 V DC, 0.625 A
Sample Stage Input:
DC Voltage and Current
24 V
0.5 A
Installation Category: CAT I (Mains isolated)
Use only with the provided power supply.
Table14.2: Power Requirements