EasyManua.ls Logo

Burkert 8175 - Temperature Adjustment; Strength of Return Signal

Burkert 8175
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
8175
ENGLISH
TEST MENU
ENTER
ENTER
4.5.3 TEMPERATURE ADJUSTMENT
The 8175 level transmitter posseses a temperature probe within the sensor. The
value from this sensor can be influenced by an offset entered within this option. To
enter a desired offset press when "T ° ADJUST" is displayed within the test
menu and enter the value.
The offset range is + or - 15°C.
The temperature adjustment remains active until another temperature
value is entered.
4.5.4 STRENGTH OF RETURN SIGNAL
This function allows the power of the ultrasonic echo recieved by the sensor to be
viewed. When the optimum signal strength is 90 dB the transmitter has a measuring
range of 7m for the remote version and 10m fo compact version. The strength of this
signal is maximum if the temperature of the gas and liquid are equal or the surface of
the fluid is calm.
This measure is useful to determine the best orientation of the transmitter
during installation. The best orientation is reached when the signal strength is
at the highest value possible.
0......9
ENTER
ENTER
Enter the temperature
offset in (°C or °F)
+ 0.0 ° C
+ 1.2 ° C
T ° ADJUST
SIGNAL
ENTER
SIMUL
4.5.3
90.00 dB
SIGNAL

Related product manuals