MAINTENANCE
5.1 STORING AND CLEANING OF THE SENSOR
Ultrasonic level sensors do not require any special maintenance, however, the bottom
of the sensor must not be covered by sediments. If necesary the sensor can be easily
cleaned using a soft bristled brush and detergent compatible with PVDF. Do not use
harsh abrasive brushes or solutions which may damage the PVDF sensor enclosure.
Ensure that the tank / pipe does not contain any fluid or pressure before
dismounting.
5.2 TROUBLE SHOOTING GUIDE
If any problems persist, please contact your local Bürkert subsidiary or
return the product with a full explanation of the problem.
This section is designed to assist with problems which may occur during installation or
operation. If in doubt, please do not hesitate to contact your local Bürkert subsidiary.
Faults Status Actions
The transmitter does not work
Transmitter connected ? No Connect the device 3.3
Fuses OK (if any ) ? No Change the fuses ----
Switch ON (if any ) ? No Set the switch ON ----
Power supply on terminal IN+ and IN- OK ? No Check the connection 3.3
Transmitter programming unavailable
Switch SW01 down (ENTER key locked) ? No Set SW01 switch down 3.3
Is configuration impossible Yes Reset to factory settings 4.5.6
Display ‘ERROR’
Display at the start-up (EEPROM failure) ? Yes Restart the device ----
Error at each start-up? Yes Return the device ----
Display after validation of the menu
(EEPROM failure) Yes Configure the device again 4.4
Failure at each validation of the menu? Yes Return the device ----
Display within the Teach-In mode ? Yes Perform the Teach-In again 4.4.6
Display ’ ---- °C’ or ’ ---- °F’
Temperature fluid between -40-80°C/-40-+176°F ? No Device out of range ----
Black connector for Pt1000 connected ? No Connect the Pt1000 ----
(Open the transmitter) connector
Blinking display of the transmitter
Blinking unit ? Yes Perform an echo reset 4.5.6
(m, cm, inch, foot, m3, l, US Gal, IMP Gal)
Blinking of the whole display ? Yes Check the connection of 3.3
connectors