GasAlertMicroClip
Sensor Configuration
2.Enter the setpoint for the CO and H
2
S sensor in the STEL Alarm
field. Proceed to STEL Interval
.
STEL Interval
STEL Interval provides protection for workers from over exposure to
high concentrations of gas, and is based on used-defined 5-15 minute
intervals. When the maximum STEL is reached, the detector alarms to
notify the worker.
a Caution
Follow all safety procedures as defined by your employer.
Enter the interval (5-15 minutes) in the STEL Interval field. The detector
is shipped with the factory default setting of 15 minutes.
Auto Zero on Startup
When enabled, the sensors automatically zero during the startup
self-tests. The Auto-Zero on Start-up option is available for the CO,
H
2
S, and LEL sensors. Not applicable to O
2
.
1. Click the checkbox of each sensor that will be auto zeroed during
startup.
The detector is shipped with the Auto-Zero on Start-up option
enabled for the CO, H
2
S, and LEL sensor.
O
2
Auto-Calibration on Startup (Automatic O
2
Calibration)
When enabled, the O
2
sensor is automatically calibrated during the
startup self-tests.
The detector is shipped with the O
2
Auto-Calibration on Startup option
enabled.
LEL By Vol CH
4
When enabled, the detector displays the LEL value as CH
4
%, assuming
a methane environment.
The LEL By Vol CH
4
option is applicable to the LEL sensor only. The
detector is shipped with the LEL by Vol CH
4
disabled.