EasyManua.ls Logo

BW Technologies GasAlertMicroClip XT - Calibrating with the IR Link; Calibration Procedure

BW Technologies GasAlertMicroClip XT
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
GasAlertMicroClip
Calibration
Calibrating with the IR Link
a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
To calibrate the detector with the IR Link, complete the following
procedure:
1. Complete steps #1-10 under Setting Span Gas Concentration
Values.
2. Click on the Device Operations tab.
3. Click
Calibrate. The detector begins calibration. Refer to Auto
Zero and Oxygen Sensor Calibration.
Calibration Procedure
a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
1. Press and hold
C. The detector performs the OFF countdown.
Continue holding
C as the detector briefly deactivates.
2. The detector then reactivates and performs the CAL countdown.
Continue holding
C until the CAL countdown is complete.
Note
If
C is not held for the entire countdown, the detector will
deactivate.
Table 12: Connecting to the IR Link
Item Description
1Detector
2 IR and charger interface
3IR Link

Table of Contents

Other manuals for BW Technologies GasAlertMicroClip XT

Related product manuals