Maintenance
To maintain the detector in good operating condition, perform the
following basic maintenance as required.
• Calibrate, bump check, and inspect the detector at regular intervals.
• Maintain an operations log of all maintenance, bump checks,
calibrations, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents,
soaps, or polishes.
• Do not immerse the detector in liquids.
Battery Cautions
a Warning
To avoid personal injury and/or property damage, adhere to
the following:
• The detector must be deactivated to charge the battery.
• Charge the battery immediately when the detector emits a low
battery alarm. Refer to Charging the Battery
.
• Charge the battery in a safe area that is free of hazardous gas in
temperatures of 32°F to 113°F (0°C to 45°C).
• Charge the battery using the BW Multi-Unit Cradle Charger or
charger adapter only. Do not use any other charging adapters.
Failure to adhere to this caution can lead to fire and/or
explosion.
• The charging adapter is voltage specific to your region. Use of
the charging adapter outside your region will damage the
charger and the detector.
• Do not calibrate the detector during or immediately after
charging the battery.
• The battery can only be replaced by the manufacturer. Failure to
adhere to this caution can lead to fire and/or explosion.
• Warning: The GasAlertMicroClip uses a lithium battery that may
present a risk of fire or chemical burn hazard if misused. Do not
disassemble, heat above 212° (100°C), or incinerate.
• Warning: Lithium polymer cells exposed to heat at 266°F
(130°C) for 10 minutes can cause fire and/or explosion.