41
GasAlertMicroClip
Calibration
When a sufficient amount of gas has been detected
(approximately 30 seconds after the gas has been applied), the
detector beeps once, flashes, and remains lit
while the detector completes the span (approximately
2minutes).
Successful Span
If the sensors have spanned successfully, the detector beeps and the
calibration procedure continues.
Unsuccessful Span
If any sensors fail the span, the following screen displays. Refer to
Calibration Troubleshooting
.
Calibration Due Date
Note
If a sensor fails calibration, the next due date for that sensor will
not reset. Refer to Calibration Troubleshooting
.
6. After calibration is complete, CAL DUE displays and all
successfully calibrated sensors automatically reset the
calibration due dates according to the calibration intervals in
Fleet Manager II.
7. The number of days that displays is when the next sensor
calibration should be performed.