EasyManua.ls Logo

BW Technologies GasAlertMicroClip XT - Clearing Gas Exposures; Gas Alarm Setpoints

BW Technologies GasAlertMicroClip XT
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GasAlertMicroClip
User Manual
32
Clearing Gas Exposures
a Caution
Follow all safety procedures as defined by your employer.
Confirm with your supervisor before clearing TWA and STEL
alarms.
To clear the TWA, STEL, and peak exposure readings, press
C when
the LCD displays RESET.
Gas Alarm Setpoints
Gas alarms are activated when detected gas concentrations are above
or below the user-defined setpoints. Gas alarms are described below.
Table 9. Gas Alarm Setpoints
Alarm Condition
Low
Toxics and combustibles: Ambient gas
level above low alarm setpoint.
Oxygen: Ambient gas level may be set
above or below 20.9% (or 20.8%).
High
Toxics and combustibles: Ambient gas
level above high alarm setpoint.
Oxygen: Ambient gas level may be set
above or below 20.9% (or 20.8%).
TWA
Toxics only: Accumulated value above the
TWA alarm setpoint.
STEL
Toxics only: Accumulated value above the
STEL alarm setpoint.
Downscale
Toxics only: If sensor reading is negative
(half of the TWA setpoint)
Multi-gas
Two or more gas alarm conditions.

Table of Contents

Other manuals for BW Technologies GasAlertMicroClip XT

Related product manuals