EasyManua.ls Logo

BW Technologies GasAlertMicroClip XT - Connecting the Gas Cylinder to the Detector

BW Technologies GasAlertMicroClip XT
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
GasAlertMicroClip
Calibration
Connecting the Gas Cylinder to the Detector
Refer to the following Figure 7., Table 11., and procedures to connect the
gas cylinder to the detector for calibration.
Note
Wind currents may cause false readings and poor calibrations.
Figure 7. Connecting the Gas Cylinder to the Detector
Table 11. Connecting the Gas Cylinder to the Detector
Read the following steps (1-7) before beginning calibration.
1. Verify the calibration gas being used matches the span
concentration value(s) that are set for the detector. Refer to
Calibration Gas in Fleet Manager II.
2. Attach a 0.5 ml/min regulator to the gas cylinder. To perform an
automated calibration, use a demand flow regulator and refer to
the MicroDock II User Manual.
3. Connect the calibration hose to the calibration cap.
4. Connect the other end of the calibration hose to the regulator on
the gas cylinder.
5. Refer to Calibration Setup
to apply gas.
6. When calibration is complete turn off gas and disconnect the
hose from the calibration cap and regulator.
7. Ensure the gas cylinder is stored according to the manufacturer’s
specifications.
Item Description
1
Calibration cap
2
Calibration hose
3
Gas cylinder with 0.5 ml/min regulator

Table of Contents

Other manuals for BW Technologies GasAlertMicroClip XT

Related product manuals