43
GasAlertMicroClip
Datalogs
Datalogs
The detector records various information that can be compiled to create
a report. The detector is capable of storing 16 hours of information (when
recording a datalog every 15 seconds). When the memory is full, the
detector replaces the oldest datalogs with the most recent datalogs
Event Logs
The detector records the 10 most recent gas alarm events. The following
information is recorded:
• Serial number of the detector
• Start time of alarm
• Type, level and duration of alarm
• Peak exposure level (ppm or %)
• Status of the sensor
Downloading Datalogs and Event Logs
The datalog and event log files can only be downloaded to a PC using an
IR Link or the MicroDock II base station. Refer to the Fleet Manager II
Operator’s Manual or MicroDock II User Manual.
Software Requirements
To create spreadsheet reports of event logs, datalogs, and bump and
calibration results, the following software applications are required:
• Fleet Manager II, and
• Microsoft Excel.