GasAlertMicroClip
User Manual
30
Alarm Screen Alarm Screen
Multi-Gas Alarm
Sequence alternating low and high alarm
siren and flash
L and gas bars flash
Vibrator alarm activates
Over Limit (OL) Alarm
Fast siren and alternating flash
L and gas bar flash
Vibrator alarm activates
OL displays
Sensor Alarm
During startup sequence Error [sensor
name] displays
During normal operation Err displays
(must be acknowledged by press C)
Confidence Beep and IntelliFlash
One beep and flash every second
Note
The detector is shipped with Confidence
Beep disabled and IntelliFlash enabled
Low Battery Alarm
Sequence of 10 rapid sirens and
alternating flashes with 7 seconds of
silence in between (continues for
15 minutes)
and L flash, LOW BAT displays,
and the vibrator alarm activates
After 15 minutes, of the Low Battery
alarm, the Automatic Shutdown Alarm
sequence begins
OFF displays before deactivating
Automatic Shutdown Alarm
Sequence of 10 rapid sirens and alternating
flashes with 1 second of silence in between
(sequence reactivates seven times)
LOW BAT and L display
Vibrator alarm activates
OFF displays before deactivating
Normal Shutdown
Sequence of two sirens and alternating flashes
Vibrator alarm activates
Countdown initiates
OFF displays