EasyManua.ls Logo

Cameron NUFLO MC-III - Automating Functions on Software Startup

Cameron NUFLO MC-III
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
MC-III™ Panel Mount Flow Analyzer Section 4
Figure 4.3—Prompt for authorizing time and date synchronization
The time difference setting used to generate a synchronize prompt is user-congurable. For more information,
see General Options, page A-1.
Automating Functions on Software Startup
When the software connects with the instrument, the Device Autorun Options screen appears (Figure 4.4).
From this screen, users can congure the instrument, download logs, or upload previously saved conguration
settings.
A checkbox at the bottom of the Device Autorun Options screen allows the user to bypass the Device Autorun
Options screen and perform a selected function each time the software connects to the instrument.
Example: To download all logs on connection, select Download all logs from the Device Autorun Options
screen and check the “Always do...” checkbox. Each time the software computer connects to the instrument,
the software will bypass the Device Autorun Options screen, open the Flow Archive screen, and download all
logs. See Section 5 for more information on ow logs.
Figure 4.4—Device Autorun Options screen

Table of Contents

Related product manuals