EasyManua.ls Logo

Cameron NUFLO MC-III - Enabling 4-20 Ma Output; 4-20 Ma Output Testing

Cameron NUFLO MC-III
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65
MC-III™ Panel Mount Flow Analyzer Section 4
Enabling 4-20 mA Output
By default, the 4-20 mA output option is disabled. To enable this feature, perform the following steps:
1. Check the “Enable” checkbox.
2. Click on the pencil icon next to the Low Flow Rate eld to launch a data-entry window (Figure 4.29) and
enter the “low” or “zero” ow rate value that will produce a 4-mA output. Adjust the decimal position,
if necessary, using the slide bar. A ow rate less than or equal to this minimum setpoint will result in an
output of 4 mA.
3. Click on the pencil icon next to the High Flow Rate eld and enter the “high” or “full-scale” ow rate
value that will produce a 20-mA output. Adjust the decimal position, if necessary, using the slide bar. A
ow rate equal to or greater than this maximum setpoint will result in an output of 20 mA.
Figure 4.29—Data-entry screen for 4-20 mA low and high values
4-20 mA Output Testing
The MC-III software allows users to test the 4-20 mA output by entering an output current value and reading
the live value generated in response.
To use the test mode, press the “Test 4-20 mA Output” button at the bottom of the 4-20 mA Output screen.
The test page will appear (Figure 4.30, page 66), and the 4-20 mA Output option will be temporarily
disabled.
To test the output, enter a value in the Output Current (mA) eld, and press “Set Output.”

Table of Contents

Related product manuals